TUBE MERCURY VAPOR .05-2MG/M3

Details:

DRAEGER Tube, Mercury Vapor 0.1/b - Order No. CH23101 Standard Measuring Range : 0.05 to 2 mg/m3
Number of Strokes (n) : 40 to 1
Time for Measurement : max. 10 min
Standard Deviation : 30 %
Colour Change : pale yellow grey >pale orange
Ambient Operating Conditions
Temperature : 0 to 40 °C
Absolute Humidity : < 20 mg H2O / L
Reaction Principle
Hg + Cul > Cu-Hg-complex
Cross Sensitivity
Free halogens cause substantial minus errors. It is impossible to measure mercury vapor in the presence of halogens. Arsine, phosphine, hydrogen sulphide, ammonia, nitrogen dioxide, sulphur dioxide and hydrazine in the TLV range do not interfere.


Product Number: 80-CH23101
Manufacturer: Draeger Safety

  $0 each